Article ; Online: Assessment of the iCare HOME2, a New Intraocular Pressure Self-Measurement Tonometer.

Kratz, Assaf / Zbidat, Rabia / Kishner, Rina / Cohen, Michal / Shalata, Walid / Goldberg, Ivan

Journal of glaucoma

2023  Volume 32, Issue 11, Page(s) 926–929

Abstract: Prcis: The iCare HOME2 tonometer generally can be considered reliable for most eyes and clinical settings, although it may produce slightly overestimated or underestimated intraocular pressure (IOP) readings in thicker and thinner corneas, respectively.! ...

Abstract Prcis: The iCare HOME2 tonometer generally can be considered reliable for most eyes and clinical settings, although it may produce slightly overestimated or underestimated intraocular pressure (IOP) readings in thicker and thinner corneas, respectively.
Purpose: To evaluate the accuracy, correlation, and analysis of differences in IOP measurements between the gold standard Goldmann applanation tonometer (GAT) and the new, self-measurement iCare HOME2 tonometer (icare).
Patients and methods: In this retrospective study, patients were randomly selected from those who attended a routine examination in our clinic. After a complete ocular examination, each patient's IOP was measured and recorded with GAT and iCare HOME2. Central corneal thickness was measured. Eyes with any corneal morbidity were excluded. Pearson correlation coefficient was used to determine the correlation between paired IOP measurements. Bland-Altman plots were graphed for the analysis of differences in IOP between the instruments.
Results: One hundred thirty-five eyes of 70 patients were included in the study. The mean IOP measured with GAT was 16.3 ± 6.5 mm Hg (range: 3-56). The mean IOP measured with iCare HOME2 was 16.5 ± 7.3 mm Hg (range: 3-55), ( P = 0.47). A strong, significant positive correlation was found for paired IOP measurements by the 2 instruments ( r = 0.94; P < 0.0001). A small systematic proportional bias was seen for the paired IOP measurements, meaning that with higher IOPs the iCare HOME2 yielded higher IOP readings than GAT, but this difference was clinically insignificant. The instrument underestimated IOPs with corneas thinner than 522 µm, whereas it overestimated IOPs when corneas were thicker than this.
Conclusion: The iCare HOME2 could be a reliable tonometer for most eyes and clinical settings. Central corneal thickness measurement is recommended in patients who use the instrument.
MeSH term(s) Humans ; Intraocular Pressure ; Ocular Hypertension/diagnosis ; Retrospective Studies ; Glaucoma/diagnosis ; Tonometry, Ocular ; Reproducibility of Results
Language English
Publishing date 2023-08-25
Publishing country United States
Document type Journal Article
ZDB-ID 913494-3
ISSN 1536-481X ; 1057-0829
ISSN (online) 1536-481X
ISSN 1057-0829
DOI 10.1097/IJG.0000000000002298
Shelf mark
Zs.A 3795: Show issues Location:
Je nach Verfügbarkeit (siehe Angabe bei Bestand)
bis Jg. 1994: Bestellungen von Artikeln über das Online-Bestellformular
Jg. 1995 - 2021: Lesesall (2.OG)
ab Jg. 2022: Lesesaal (EG)
Database MEDical Literature Analysis and Retrieval System OnLINE

More links

Kategorien

To top