Article ; Online: Technical Note: Performance evaluation of a small-animal PET/CT system based on NEMA NU 4-2008 standards.
2021 Volume 48, Issue 9, Page(s) 5272–5282
Abstract: Purpose: The Metis: Methods: Energy resolution, spatial resolution, sensitivity, scatter fraction (SF), noise-equivalent count rate (NECR), and image quality (IQ) characteristics were measured. A micro Derenzo phantom experiment was performed to ... ...
Abstract | Purpose: The Metis Methods: Energy resolution, spatial resolution, sensitivity, scatter fraction (SF), noise-equivalent count rate (NECR), and image quality (IQ) characteristics were measured. A micro Derenzo phantom experiment was performed to evaluate the spatial resolution using three-dimensional ordered-subsets expectation maximization (3D-OSEM) and maximum likelihood expectation maximization (MLEM) reconstructed images. In addition, the CT imaging agent Ioverol 350 was mixed with fluorine-18 ( Results: The mean energy resolution of the system was 15.3%. The tangential resolution was 0.82 mm full-width half-maximum (FWHM) at the center of the field of the view (FOV), and the radial and axial resolution were generally lower than 2.0 mm FWHM. The spatial resolution was significantly improved when using 3D-OSEM, especially the axial FWHM could be improved by up to about 57%. The system absolute sensitivity was 7.7% and 6.8% for an energy window of 200-750 and 350-750 keV respectively. The scatter fraction was 8.2% and 12.1% for the mouse- and rat-like phantom respectively. The peak NECR was 1343.72 kcps at 69 MBq and 640.32 kcps at 53 MBq for the mouse- and rat-like phantom respectively. The 1-mm fillable rod in the IQ phantom can be clearly observed. We can identify the 0.6-mm aperture of the micro Derenzo phantom image clearly using 3D-OSEM (10 subsets, 5 iterations). We also performed the fusion of the PET and CT images of the mouse and the brain imaging of the rat. Conclusions: The results show that the system has the characteristics of high-resolution, high-sensitivity, and excellent IQ and is suitable for rodent imaging-based research. |
|||||
---|---|---|---|---|---|---|
MeSH term(s) | Animals ; Fluorodeoxyglucose F18 ; Mice ; Phantoms, Imaging ; Positron Emission Tomography Computed Tomography ; Positron-Emission Tomography ; Rats ; Reference Standards | |||||
Chemical Substances | Fluorodeoxyglucose F18 (0Z5B2CJX4D) | |||||
Language | English | |||||
Publishing date | 2021-07-26 | |||||
Publishing country | United States | |||||
Document type | Journal Article | |||||
ZDB-ID | 188780-4 | |||||
ISSN | 2473-4209 ; 0094-2405 | |||||
ISSN (online) | 2473-4209 | |||||
ISSN | 0094-2405 | |||||
DOI | 10.1002/mp.15088 | |||||
Shelf mark |
|
|||||
Database | MEDical Literature Analysis and Retrieval System OnLINE |
Full text online
More links
Kategorien
In stock of ZB MED Cologne/Königswinter
Zs.A 1210: Show issues | Location: Je nach Verfügbarkeit (siehe Angabe bei Bestand) bis Jg. 1994: Bestellungen von Artikeln über das Online-Bestellformular Jg. 1995 - 2021: Lesesall (1.OG) ab Jg. 2022: Lesesaal (EG) |
Order via subito
This service is chargeable due to the Delivery terms set by subito. Orders including an article and supplementary material will be classified as separate orders. In these cases, fees will be demanded for each order.