Article ; Online: [Consideration of Lens of the Eye Protection during Mammography with the Eyelid and the Face Guard].

Sekizawa, Honami / Igarashi, Takayuki / Yamamoto, Emiko / Okumura, Kenichirou / Imai, Hirohisa / Masuda, Yoshitada

Nihon Hoshasen Gijutsu Gakkai zasshi

2021  Volume 77, Issue 4, Page(s) 358–364

Abstract: Purpose: In 2011, the International Commissionon Radiological Protection (ICRP) recommended reducing the threshold dose for the lens. Therefore, it is important to reduce the lens exposure dose in medical exposures. In a cranio-caudal (CC) view of ... ...

Abstract Purpose: In 2011, the International Commissionon Radiological Protection (ICRP) recommended reducing the threshold dose for the lens. Therefore, it is important to reduce the lens exposure dose in medical exposures. In a cranio-caudal (CC) view of mammography, the patient' s lens receives scattered radiation. In this study, we investigated scatter dose around the lens during mammography and reviewed the simple and easy protection methods of the lens.
Methods: Optically stimulated luminescence (OSL) dosimeters were placed in front of the device to obtain scattered radiation intensity distribution. The human phantom was placed in the same way as the CC positioning, and BR-12 phantoms with a thickness of 40 mm was placed on the FPD. Then, the scatter dose around the lens was measured using an OSL dosimeter. In order to confirm the change in the scatter radiation dose by the face guard (FG) and eyelid, we measured and compared under the same conditions the presence of FG and adipose tissue about 1 mm thick assuming the eyelid.
Results: Scatter radiation intensity decreased around the FG. When the FG was installed, the scatter dose was reduced about 33%, and when the adipose tissue was pasted on the OSL dosimeter, the scatter dose was reduced about 29%.
Conclusion: This study suggested that eye closure during mammography was effective in reducing lens exposure. In the future, we would like to expect further protective effects by increasing the thickness of FG and reviewing the materials.
MeSH term(s) Eyelids ; Humans ; Lens, Crystalline ; Mammography ; Radiation Dosage ; Radiation Protection
Language Japanese
Publishing date 2021-04-21
Publishing country Japan
Document type Journal Article
ZDB-ID 2269092-X
ISSN 1881-4883 ; 0369-4305
ISSN (online) 1881-4883
ISSN 0369-4305
DOI 10.6009/jjrt.2021_JSRT_77.4.358
Database MEDical Literature Analysis and Retrieval System OnLINE

More links

Kategorien

To top