Artikel ; Online: Using the Seraph® 100 Microbind® Affinity blood filter under slow flow conditions through 18 G and 16 G central lines.
The journal of vascular access
2024 , Seite(n) 11297298241234073
Abstract: Introduction: The Seraph® 100 Microbind® Affinity blood filter (Seraph® 100) has been in use since 2019 for the treatment of fulminant or difficult to treat blood stream infections as an adjunct to pharmacotherapy. In 2020 the device received emergency ... ...
Abstract | Introduction: The Seraph® 100 Microbind® Affinity blood filter (Seraph® 100) has been in use since 2019 for the treatment of fulminant or difficult to treat blood stream infections as an adjunct to pharmacotherapy. In 2020 the device received emergency use authorization by the US Food and Drug Administration for the treatment of critically ill COVID-19 patients with confirmed or imminent respiratory failure. Results of an international registry showed that the Seraph® 100 was operated under blood flow rates of 100-350 mL/min. As those conditions require a large bore central line, a dialysis catheter is currently considered indispensable to operate the Seraph® 100. The use of smaller catheter lumina has neither been evaluated in vitro nor in vivo. Methods: In vitro pressure data before and after the Seraph® 100 at various blood pump rates (prepump line 16 G, postpump line 18 G) with saline and human plasma were obtained. Further, anecdotal flow and pressure data of two patients treated with the Seraph® 100 for a COVID-19 infection are reported. Results: At a pump speed of 50 mL/min pre-Seraph® pressure using saline was -70 [-70 to -60] mm Hg. In comparison, using plasma pre-Seraph® pressure was lower at -120 [-120 to -105] mm Hg; Conclusion: Seraph® 100 blood flow rate of 50 mL/min may be achieved using low flow vascular access, allowing to treat a blood volume 72 L in 24 h. |
---|---|
Sprache | Englisch |
Erscheinungsdatum | 2024-02-28 |
Erscheinungsland | United States |
Dokumenttyp | Journal Article |
ZDB-ID | 2252820-9 |
ISSN | 1724-6032 ; 1129-7298 |
ISSN (online) | 1724-6032 |
ISSN | 1129-7298 |
DOI | 10.1177/11297298241234073 |
Datenquelle | MEDical Literature Analysis and Retrieval System OnLINE |
Zusatzmaterialien
Kategorien
Verfügbar in ZB MED Köln/Königswinter
Zs.A 6414: Hefte anzeigen | Standort: Je nach Verfügbarkeit (siehe Angabe bei Bestand) bis Jg. 2021: Bestellungen von Artikeln über das Online-Bestellformular ab Jg. 2022: Lesesaal (EG) |
Über subito bestellen
Dieser Service ist kostenpflichtig (siehe Lieferbedingungen von subito). Bestellungen, die einen Artikel nebst Supplementary Material umfassen, werden grundsätzlich wie mehrfache Bestellungen bearbeitet. Gebühren fallen in diesen Fällen für jede einzelne Bestellung an.